Fluorescence Spectroscopy
Reference: Harris, pp. 525-533; Skoog/Leary pp.
130-138, 355-374
Fluorescence
is the emission of light by a molecule, which has absorbed radiant energy; the radiation is emitted at a longer
wavelength (lower energy) than the incident absorbed energy. The following figure illustrates
absorption transitions from the ground state to various vibrationally excited
states of the upper electronic energy on the left.
Refer
to Figure 15-1, pp. 357
Skoog/Holler/Nieman
So represents the singlet ground state (all electrons are
paired). S1 refers
to the lowest singlet excited state (all electrons are paired). Occasionally, excitation of an electron
to a higher energy level can result in a triplet state T1 (unpaired
electrons).
Electrons in the excited state (with a lifetime of about 10-8
s) will drop to the lowest vibrational energy level due to collisions via
vibrational deactivation or relaxation.
Electrons can now return to the ground state by internal conversion,
emitting heat through vibrational relaxation. This is accomplished by returning to the
ground state via various vibrational energy levels. They can also return to various
vibrational energy levels of So at a longer wavelength (lower energy)
by fluorescence. The longer
wavelength is a consequence of loss of energy during vibrational
deactivation. In certain cases,
electrons cross over to the triplet state T1 via intersystem
crossing. When they return to the
various vibrational levels of So the light emitted is known as
phosphorescence. Phosphorescence, a
result of a "forbidden transition", has a longer lifetime (10-4 s
---> hours) than does fluorescence.
Saturated molecules and molecules with only one double bond do not
exhibit significant fluorescence.
Molecules with at least one aromatic ring or multiple conjugated double
bonds are prone to having fluorescence spectra in the visible region. Substituents such as -OH,
-OCH3, and -NH2 which are electron donating groups can
enhance fluorescence.
Consult your textbook for a discussion of the relation between
fluorescence intensity and concentration.
A plot of the fluorescence intensity vs. concentration should be linear
(at low concentrations). Reduction
in intensity of fluorescence can be due to specific effects of constituents of
the solution itself. The term
quenching is used to describe any such reduction in intensity. Types of quenching include concentration quenching (a decrease in
the fluorescence per unit concentration as the concentration is increased), also
referred to as an inner filter effect, collisional quenching and chemical quenching. Concentration quenching results from
excessive absorption of either primary or fluorescent radiation by the
solution. Collisional quenching may
be caused by nonradiative loss of energy from the excited molecules, and the
quenching agent (such as oxygen) may facilitate conversion of the molecules from
the excited singlet to triplet level.
Chemical quenching is due to actual changes in the chemical nature of the
fluorescent substance such as conversion of a weak acid to its anion with
increasing pH. Aniline is an
example. It fluoresces as the
molecule between pH 5 and pH 13, below pH 5 it exists as the cation, and above
pH 13 it exists as the anion; both
do not fluoresce.
In order to separate the emitted radiation from the incident beam (the
source), fluorescence measurements are made at right angles to the incident
beam. This is possible because
fluorescence is emitted in all directions, but the incident radiation passes
straight through the cell. Think of what would happen if the geometry of a
fluorescence instrument were in a straight line (180o) rather than
90o as is the case.
Incident light from the source would travel through the excitation
monochromator, through the sample cell, straight through the emission
monochromator and directly to the photomultiplier tube. The detector would sense mostly light
emanating from the source with a small amount of fluorescence from the
sample. It is much easier to detect
fluorescence emission from the sample in the absence of light from the
source. The 90o geometry
makes it possible to reduce as much as possible light from the source from
reaching the detector.
The source of a spectrofluorometer is usually a high intensity Xe arc
lamp which yields a continuous emission from 200 ---> 800 nm. The radiation passes through a
(excitation) monochromator to allow a certain range of wavelengths to strike the
sample. The fluorescent light
generated by the sample passes through a second (emission) monochromator aligned
90o to the source. The
fluorescent light then strikes a photomultiplier tube, which transforms the
light signal into electrical energy.
An emission (fluorescence) spectrum is obtained by holding the excitation
monochromator at a fixed wavelength and scanning the emission
monochromator. The excitation
monochromator is set by determining a wavelength at which the sample absorbs
radiation. An excitation
(fluorescence) spectrum is obtained in the opposite way. That is, by scanning the excitation
monochromator while holding the emission monochromator at a fixed
wavelength. The emission
monochromator is set by determining a wavelength at which the sample emits
radiation. The mirror image
relationship between emission and excitation spectra is due to the fact that the
vibrational levels in the ground and excited states have nearly the same
spacing, and the molecular orbital symmetries do not change. Assuming that all of the molecules are
in the ground state before excitation, the least energy absorbed in the
excitation process (i.e. longest wavelength) equals the greatest energy
transition in the fluorescence process (i.e. shortest
wavelength).